Quality control of noninvasive voltage meters in X-ray equipment, diagnostic level

Controle de qualidade de medidores de tensão não-invasivos em equipamento de raios X, nível diagnóstico

Renato Assenci Ros, Linda V.E. Caldas



ABSTRACT

We present the quality control tests performed on a monophasic X-ray equipment and on two noninvasive monitoring systems. The results were compared with the values obtained using an invasive monitoring system and a hyperpure germanium detector, providing correction factors for utilization of the noninvasive voltage meters in the quality control program of X-ray equipment. We concluded that there is a need for periodic calibration and control.

Keywords: X-radiation, Quality control, Noninvasive systems

RESUMO

Este trabalho apresenta os testes de controle de qualidade feitos em um equipamento de raios X monofásico e em dois sistemas de monitoração não-invasivos. Os resultados foram comparados com os valores apresentados por um sistema de monitoração invasivo e por um detector de germânio hiperpuro, fornecendo fatores de correção para utilização dos medidores de tensão não-invasivos no programa de controle de qualidade de equipamentos de raios X, mostrando a necessidade de calibração e controle periódico.

Palavras-chave: Radiação X, Controle de qualidade, Sistemas não-invasivos


THE CONTENT OF THIS ARTICLE IS NOT AVAILABLE FOR THIS FORMAT.


Creative Commons License
This work is licensed under an Attribution 4.0 International License (CC BY 4.0), effective June 9, 2022. Previously, the journal was licensed under a Creative Commons Attribution - Non-Commercial - Share Alike 4.0 International License.

Site Map



  • SPONSORED BY

Av.Paulista, 37 - 7° andar - Conj. 71 - CEP 01311-902 - São Paulo - SP - Brazil - Phone: (11) 3372-4554 - Fax: (11) 3372-4554

© All rights reserved 2025 - Radiologia Brasileira